266
edits
Changes
Jump to navigation
Jump to search
Part 2
The right side of the unit has banana plugs for connecting the device under test (DUT). If you're not used to so many terminals, this is because this instrument can do 4-wire resistance, which a normal handheld meter cannot do. The two connectors in the middle column are used if you're used to a normal handheld multimeter.
=== Changing Settings ===
In order to change a setting, type the number you want the setting to be on the keypad. Then, press the "STORE" button, then press the keypad button with the white text that corresponds to the setting you want to change. Here are some common settings:
* '''N CYC INT''': Number of powerline integration cycles. This unit supports up to 100 powerline integration cycles. The higher the number, the more accurate the measurement but the longer the measurement time. The default is 10.
* '''N DIG DISP''': Number of visible digits. This DMM supports 2, 3, 4, 5, or 6 visible digits. You can change this setting to account for the uncertainty of the measurement you are performing. The default is 5.